Different specification of gel card
There are kinds of gel card and each one
have the different use and function.Below is the normal kind coombs gel card which normal use in hospital and laboratory.
1.The gel shall allow a free passage for
the unagglutinated red cells to pellet at the bottom of the microtubes when
centrifuged to exhibit a distinct negative reaction.
2.The gel shall be able to keep strongly
agglutinated red cells completely remaining the top to exhibit a 4+ reaction.
3.Lesser reactions shall be proportionately
obstructed to leave a trail of agglutinated cell trapped in the gel enabling
the reaction to be graded.
4.The gel shall be able to hold the
reactions in place for later review for at least a week.
5.Inversion of the microtubes shall not
cause the contents to be dislodged which may result in discrepancy between
initial reactions obtained and reactions reviewed later.
6.The gel card must be compatible and used
on the existing equipment: Gel Card Centrifuge & Incubator.
7.The gel card must be compatible and used
on the current existing reagents: Antibody Screening Cells and Antibody
Identification panel cells).
8.Test sample can be used either on EDTA
blood or drawn plain tubes.
9.Vendor must provide 1 unit of floor standing
Fully Automated Immunohematology System for the gel card.